VHS-10 &
VHS-400
Diffusion
Pumps
Instruction Manual
Varian Vacuum Technologies
121 Hartwell Avenue
Lexington, MA 02421-3133
(781) 861-7200
(800) 8-VARIAN
Manual No. 6999-01-023
Revision H
October 2000
Vedere la pagina 0
1 2 3 4 5 6 ... 35 36

Sommario

Pagina 1 - Diffusion

VHS-10 &VHS-400DiffusionPumpsInstruction ManualVarian Vacuum Technologies121 Hartwell AvenueLexington, MA 02421-3133(781) 861-7200(800) 8-VARIANMa

Pagina 2 - Sales and Service Offices

viii. Cold Coolant – Liquid nitrogen, a cryogenic liquid, is used in traps. If it is splashed on body tissuesor eyes, it can cause severe frostbite or

Pagina 3

viiiWARNING_____________High voltages (up to 480 V) can kill. Always break the primary circuit tothe power supply before starting to work on the heate

Pagina 4

ixNOTE________Always dispose of used or dirty oil properly and in compliance with alllocal, state, and federal environmental laws and regulations.CAUT

Pagina 6

Table of ContentsSection PageSafety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Pagina 7

1-1SECTION ISPECIFICATIONS AND INSTALLATIONBefore unpacking and installing the VHS-10 or the VHS-400 Diffusion Pumps, the user should thor-oughly fami

Pagina 8 - Prohibited Action Result

Specifications and Installation1-2Table 1-1. VHS-10 Pump Specifications (Cont'd)Foreline Baffle Stacked, self-aligning half moonCold Cap Nickel-p

Pagina 9

Specifications and Installation1-3Table 1-2. VHS-400 Pump Specifications (Cont'd)Pressure drop across coils 6 psi Backing Pump Size Recommended 3

Pagina 10

Specifications and Installation1-41-2-2 Pump Oil Installationa The recommended oil charge for the VHS-10 and the VHS-400 diffusion pumps is 1000 cc.b

Pagina 11

Specifications and Installation1-5b The exit water temperature should not exceed 120°F.c The outlet or discharge (nearest the foreline) should be conn

Pagina 12

United StatesVarian Vacuum Products121 Hartwell AvenueLexington, MA 02421USA Tel: (781) 861 7200Fax: (781) 860 5437Toll Free: 1 (800) 882 7426Argentin

Pagina 13

Specifications and Installation1-6b Connect the leads of the thermal switch in series with the coil of the heater power supply. Inthe event of overtem

Pagina 14 - Table Title Page

2-1SECTION IIOPERATIONCAUTION________Damage to PumpDo not turn on the heater without fluid in the pump. This may ruin theheaters and damage the pump.D

Pagina 15

Operation2-22-2 SHUTDOWN PROCEDUREa Turn off the power to the diffusion pump. Continue to back the diffusion pump with the appro-priate mechanical pum

Pagina 16

3-1SECTION IIIMAINTENANCE AND SERVICE3-1 GENERALDiffusion pumps generally require little attention when operated correctly. It is advisable to perform

Pagina 17

Maintenance and Servicef Thoroughly clean the diffusion pump body interior and the jet assembly using acetonefollowed by an isopropyl alcohol rinse. D

Pagina 18

Maintenance and Service3-3Figure 3-1 Jet Assembly, VHS-10 and VHS-400 Pumps

Pagina 19

Maintenance and Service3-4To install the jet assembly, proceed as follows.a Place the splash baffle in the bottom of the pump. Check that it is locate

Pagina 20

Maintenance and Service3-5d Loosen and remove the fill plug located at the top of the sight glass assembly.e Add pump fluid until the fluid level is e

Pagina 21 - ________

Maintenance and Service3-6

Pagina 22

4-1SECTION IVTROUBLESHOOTING4-1 LEAKAGEAnalysis of general operational experience with diffusion pumps indicates that certain locations aremore prone

Pagina 23

WARRANTYProducts manufactured by Seller are warranted against defects in materials and workmanship fortwelve (12) months from date of shipment thereof

Pagina 24

Troubleshooting4-2Table 4-1 TroubleshootingFault Probable Cause Corrective ActionPoor system pressure Leaks in system, virtual or real Locate and repa

Pagina 25 - Maintenance and Service

5-1SECTION VREPLACEMENT PARTSThe following is a list of replacement parts for the Varian VHS 10 and VHS 400 Diffusion Pumps.Table 5-1 Replacement Part

Pagina 26

Replacement Parts5-2Table 5-1 Replacement Parts (Cont’d)Part Number Description K0377169 O-ring kit; includes:(VHS-400 only) 1 Butyl inlet O-ring, Par

Pagina 27

Replacement Parts5-3Figure 5-1 Wiring Diagram,VHS-400 Diffusion Pump

Pagina 28

Replacement Parts5-4VHS-10 Pump VHS-400 PumpVHS-10 VHS-400ASA ISO ConFlat ASA ISOin mm in mm in mm in mm in mmA Height 31.00 787 31.00 787 31.00 787 3

Pagina 29

Replacement Parts5-5Figure 5-3 VHS-10 Pump Performance CurvesFigure 5-4 VHS-400 Pump Performance CurvesNOTE________Speed curves were generated accordi

Pagina 30 - Table 4-1 Troubleshooting

Replacement Parts5-6Old Style – Typical New Style – TypicalFigure 5-5 Fill and Drain Assemblies

Pagina 31 - Table 5-1 Replacement Parts

iOPERATION AND MAINTENANCE OF THIS EQUIPMENT INVOLVES SERIOUS RISK. IT IS THERESPONSIBILITY OF THE USER TO MAINTAIN SAFE OPERATING CONDITIONS AT ALL T

Pagina 32

iiSAFE INCORPORATION OF VARIAN DIFFUSION PUMPS INTO VACUUM SYSTEMS IS THERESPONSIBILITY OF THE SYSTEMS DESIGNER. TAKE APPROPRIATE ACTION THROUGHREDUND

Pagina 33 - Replacement Parts

iiiTHE INSTALLATION, OPERATION, AND/OR SERVICING OR DIFFUSION PUMPS INVOLVES ONEOR MORE OF THE FOLLOWING HAZARDS, ANY ONE OF WHICH, IN THE ABSENCE OF

Pagina 34 - VHS-10 Pump VHS-400 Pump

ivHazard Suggested Corrective ActionExplosion Integrate pressure relief valves in all systems using pumps 10 inches or larger in diameter.Do not use o

Pagina 35

vProhibited Action Explosion-Causing ConditionOpen drain or fill plug while pump is Overpressure plus strong oxidizerunder vacuum, especially when it

Pagina 36

vid. Poisonous and/or Corrosive Compounds – When pumping poisonous, reactive, and/or corrosivegas, vapors, or chemicals, even proper operation and reg